| Oxygold-G | Oxygold-B | Oxyferm | Oxyferm-XL | Oxysens | Oxysens-W | |
| Limite détection | 2 ppb | 10 ppb | 10 ppb | 10 ppb | 40 ppb | 40 ppb |
| Tension de polarisation | - 670 mV | 0 mV | - 670 mV | - 670 mV | - 670 mV | - 670 mV |
| Courant dans air | 180-400 nA | 180-400 nA | 40-80 nA | 40-80 nA | 40-80 nA | 40-80 nA |
| Temps de réponse t98 | 30-60 sec. | 30-60 sec. | 30-60 sec. | 30-60 sec. | 60 sec. | 60 sec. |
| Débit min. | 0.1 m/sec | 0.1 m/sec | 0.02 m/sec | 0.02 m/sec | 0.02 m/sec | 0.02 m/sec |
| Température de mesure | 0-130 °C | 0-100 °C | 0-130 °C | 0-130 °C | 0-60 °C | 0-60 °C |
| Pression max. adm | 0-12 bar | 0-12 bar | 0-4 bar | 0-4 bar | 0-4 bar | 0-4 bar |
| Raccord process | PG 13.5 | PG 13.5 | PG 13.5 | G 1 ¼ " | PG 13.5 | PG 13.5 |
| Connection électrique | VP 6.0 | VP 6.0 | std 4 pôles | std 4 pôles | Câble 5 m | Câble 5 m |
| Capteur de température | NTC 22 KW | NTC 22 KW | NTC 22 KW | NTC 22 KW | NTC 22 KW | NTC 30 KW |